Lab Q

ASTM Type II: Lab-Q Smart

ASTM Type II: Lab-Q Smart

Features

INDION Lab-Q smart (Type II) water maker produces pure water which is essential for creating chemical reagents, microbiological buffers and media that are consistently pure. It is an excellent source of water for glassware washing and as feed water for laboratory equipment such as autoclaves, laboratory dishwashers and water baths. It is also a recommended source of pre-treated water for user in Type I ultrapure systems. It produces water with very low total organic carbon (TOC) content and conforms to ISO 3696 and corresponding ASTM standards. Portable tap water can be directly connected to INDION Lab-Q Smart to produce Type II pure water. It is easy to install and has several built-in safety features for reliable and consistent Type II water supply. For feed water with higher level of TDS i.e. > 500 ppm a customised pre-treatment module can be designed prior to INDION Lab-Q Smart.

  • Features
  • Online resistivity meter – displays product water quality for continuous monitoring
  • Automatic flush device (AFD) – ensure periodic cleaning of the system
  • Dry run protection – ensures protection of the pump when inlet water is unavailable
  • Easy accessibility of components – all cartridges and filters are easily accessible and no special tool is required to replace them.
  • Simple installation – simple to install, operate and maintain with a clear indication of water purity
  • Easy to dispense pure water
  • Electrolytic auto sanitizer – ensures periodic sanitization for upstream purifying filters
  • Optional filtration – filter can be attached at the point of dispensing spout to ensure compliance to microbial reduction values
  • Optional flexible dispensing – remote water dispenser offers dispensing at various angles and height
  • Advantages
  • Works on potable water as the feed
  • Compact and portable counter tap model ensures optional use of space
  • Low operating cost
  • Requires minimum maintenance
  • Applications
  • Flame spectrometry
  • Electrochemistry
  • Sample dilution
  • Radioimmunoassay
  • Electrophoresis
  • General chemistry
  • Feed for Type I water maker
  • Feed for clinical analyzers
  • Preparation of media and buffer
  • Autoclave and water bath

Feed Water Specifications

Configuration10 LPH50 LPH
TDS500 ppm max.
Fe, MnNil
Free chlorineNil
Silica< 20 ppm
Inlet pressure0.5 – 2 .0 Kg/cm²Min 0.5 kg/cm²
Max 2.0 kg/cm²

Product Specifications

Configuration10 LPH50 LPH
Dimensions
W x D x H (in mm)
450 x 250 x 500390 x 740 x 720
Weight13 kgs40 kgs
Power Requirement230 VAC ± 10%
Power Consumption40 Watts250 Watts

Performance Parameters

Flow Rate10 LPH50 LPH
Resistivity (Grade II)> 10 MΩ.cm
Conductivity (Grade II)< 0.1 μS/cm
Endotoxin level*< 0.25 EU/ml
Particle > 0.22 μm*< 1 per ml
Bacteria*< 0.1 cfu/ml< 10 cfu/100ml
Total organic carbon (TOC)*< 30 ppb

Note: * Performance parameters achieved when water dispensed through final point of use 0.22 μ

* TOC values achieved with feed water of less than 2000 ppb